What is Delivered Dose Uniformity?

Delivered Dose Uniformity, or DDU, is used to assess how consistently an inhaler or nasal drug product delivers the intended dose from the device. It is an important part of OINDP testing because the amount of active drug emitted from the device needs to remain consistent across repeated actuations, between devices and throughout the labelled life of the product.
A typical DDU test involves actuating the device into a dose collection apparatus, collecting the emitted dose and then recovering and analysing the sample. The exact test set-up depends on the device type, which is why different approaches are used for MDIs, DPIs, nebulisers, soft mist inhalers and nasal products.

Delivered Dose Uniformity Testing by OINDP Type

Select your OINDP type to explore the most suitable Delivered Dose Uniformity (DDU) testing set-up for your device. Copley supports DDU testing for MDIs, DPIs, nebulisers, SMIs and nasal products, with equipment designed to support reliable dose delivery assessment.

DDU of Metered-Dose Inhalers (MDIs)
DDU testing for metered-dose inhalers measures the consistency of the emitted dose from the device during actuation. Because MDI aerosolisation is driven by the propellant rather than the patient’s inspiratory effort, standard MDI DDU testing is typically carried out at a fixed test flow rate. Copley provides MDI DDU testing solutions to support repeatable device actuation, dose collection and sample recovery, including equipment designed for testing MDIs with add-on devices such as spacers and valved holding chambers.
View MDI DDU Test Set-Up
DDU of Dry Powder Inhalers (DPIs)
DDU testing for dry powder inhalers measures the dose emitted from the device under controlled airflow conditions. As most DPIs are passive, breath-actuated devices, the patient’s inspiratory effort, device resistance and airtest flow rate all play an important role in dose delivery. Copley provides a comprehensive range of DPI DDU testing equipment, including dose collection apparatus, critical flow controllers and supporting systems designed to help laboratories establish reliable, repeatable test conditions in line with relevant pharmacopoeial and regulatory expectations.
View DPI DDU Test Set-Up
DDU of Nebulisers
DDU testing for nebulisers is used to determine the amount of drug delivered during a defined treatment period. Rather than assessing one single actuation, nebuliser DDU testing considers dose delivery over time, often using breathing simulation to create more representative test conditions. Copley provides nebuliser DDU testing equipment to support delivered dose, active substance delivery rate and total active substance delivered testing in line with relevant pharmacopoeial requirements.
View Nebuliser DDU Test Set-Up
DDU of Soft Mist Inhalers (SMIs)
DDU testing for soft mist inhalers assesses the consistency of dose delivery from devices that generate a slow-moving aerosol cloud. As soft mist inhalers are active, aqueous-based devices, testing is often similar to MDI DDU testing and is typically carried out at a constant test flow rate. Copley provides SMI DDU testing equipment to support delivered dose collection and recovery across repeated actuations and device life.
View SMI DDU Test Set-Up
DDU of Nasal Products
DDU testing for nasal products measures the consistency of the dose emitted from nasal sprays, nasal aerosols and nasal powders. This helps laboratories assess dose delivery from devices designed for local nasal delivery or wider systemic absorption. Copley provides nasal product DDU testing equipment and supporting set-ups for different nasal drug delivery formats, helping laboratories select the right approach for emitted testing and sample recovery.
View Nasal Product DDU Test Set-Up

Delivered Dose Uniformity FAQs

What is Delivered Dose Uniformity?

Delivered Dose Uniformity is a measure of how consistently an inhaler or nasal drug product delivers the intended dose from the device. It is used in OINDP testing to assess whether dose delivery is consistent between batches, that the number of dose deliveries are greater than or equal to label claim and that dose delivery remains consistent across the life of the product in the case of multi-dose devices.

What is a DDU test?

A DDU test measures the amount of active drug delivered or emitted from an inhaler or nasal device under controlled test conditions. The dose is usually collected using a sampling apparatus before being recovered and analysed.

Is delivered dose the same as emitted dose?

Delivered dose and emitted dose are closely related, but they are not always used in exactly the same way. Delivered dose usually refers to the amount of drug available to the user, while emitted dose is often used to describe the amount of drug released from the device during testing.

Which OINDPs require DDU testing?

DDU testing is relevant to all orally inhaled and nasal drug products, including metered-dose inhalers, dry powder inhalers, nebulisers, soft mist inhalers, nasal sprays, nasal aerosols and nasal powders.

Why is DDU testing important?

DDU testing is important because it helps confirm whether an inhaler or nasal drug product can deliver a consistent dose. Consistent dose delivery supports product quality, patient safety and therapeutic efficacy, while also helping laboratories build reliable methods and generate data that meets regulatory expectations.