APSD of Metered Dose Inhalers (MDIs)

The APSD testing of MDIs is typically performed at a flow rate of 28.3 L/min when using an ACI or 30 L/min when using an NGI. For Breath Actuated MDIs (BAIs) a Breath Actuation Controller may also be used to generate a time delay. However, it must be noted that when add-on devices such as spacers or valved holding chambers (VHCs) are used, the APSD characteristics may be substantially altered from what is emitted when the MDI is used alone and must be assessed appropriately.

Breath-Actuated MDIs (BAIs)

The Breath Actuation Controller BAC 100i can also be used for the testing of Breath-Actuated (or Breath-Operated) MDIs. In this case, the BAC 100i is used to initiate the flow, simultaneously triggering the breath-actuated inhaler.

Semi-Automation Tools

Improve efficiency
Reduce variability
Eliminate handling error
Increase testing capacity

How Does a Cascade Impactor Work?

Play Video

Related Services

Related News & Resources

New e-Training service from Copley for pharmaceutical testing

29th January 2021

Find out more

New, state-of-the-art vision system central to Copley Scientific’s cascade impactor validation service

24th July 2007

Find out more