Contact Us
Contact Us
The established NGI platform for routine APSD testing
Aerodynamic particle size is a critical quality attribute for OINDPs, influencing where drug particles deposit in the respiratory tract. Cascade impaction is widely used for APSD measurement owing to its ability to separate the delivered dose into aerodynamic size fractions for HPLC analysis, enabling drug-specific particle size distribution data rather than a non-specific measurement of the whole aerosol.
Developed specifically to meet the needs of pharmaceutical inhaler testing, the NGI has become an established platform for APSD measurement in laboratories worldwide. It is specified in the European Pharmacopoeia as Apparatus E (Ph. Eur. 2.9.18) and in the United States Pharmacopeia as Apparatus 5 and 6 (USP <601>), and operates at flow rates from 15 to 100 L/min.
It was designed to overcome practical limitations of earlier impactors, with a horizontal layout and removable collection cups that make dose recovery quicker and easier to automate. It is now one of the most widely used cascade impactors for MDI, DPI, SMI, nebuliser and nasal product testing.
Why choose Copley for your NGI?
Choosing an NGI is not just about selecting the impactor itself. Reliable APSD measurement depends on how the full method is configured, controlled, maintained and supported over time.
Copley brings together precision NGI instrumentation with the practical inhaler testing expertise needed to apply it effectively across development, quality control and regulatory workflows. This gives laboratories access to an established NGI platform, proven accessories, stage mensuration, servicing and ongoing technical support to help maintain confidence in their APSD data.
- NGIs designed and manufactured by Copley in Nottingham, UK
- NGI stage mensuration, servicing and calibration support
- Inhaler Testing Academy® training on cascade impaction method set-up and data analysis
Key features: NGI
Meets Ph. Eur. 2.9.18 (Apparatus E) and USP <601> (Apparatus 5 and 6) specifications
Low inter-stage wall losses for efficient drug recovery and mass balance
Seven stages plus a Micro-Orifice Collector (MOC), giving high-resolution size fractionation from 15 to 100 L/min
Excellent stage efficiency, accuracy and reproducibility supporting efficient sample recovery and day-to-day operation
Electrically conductive design to help minimise electrostatic effects
Removable cup tray for fast, consistent dose recovery and easy automation
NGI specifications
| Pharmacopoeial references | Ph. Eur. 2.9.18 Apparatus E; USP <601> Apparatus 5 (with preseparator) and Apparatus 6 (without preseparator) |
|---|---|
| Stages | Seven impaction stages plus a Micro-Orifice Collector (MOC) |
| Flow rate range | 15 to 100 L/min |
| Stage cut-off range | 0.24 to 14.1 µm (100 to 15 L/min) |
| Compatible accessories | Induction port, preseparator, NGI dissolution cups, NGI Cooler™, NGI Tilting Platform, Impactor Coater IC 200i, Sample Recovery System™ SRS 100i |
| Calibration | Archival calibration data; annual stage mensuration recommended |
How the Next Generation Impactor works
APSD measurement is a key test used to understand how an inhaled or nasal drug product performs. It measures the size distribution of aerosolised drug particles according to their aerodynamic behaviour, rather than their geometric size alone.
For OINDPs, APSD is a critical quality attribute. Particles that are too large may deposit in the mouth, throat or device-related components, while smaller particles may be more likely to reach the lower airways. By characterising the delivered dose by aerodynamic size, APSD testing helps laboratories assess whether a product is delivering drug to the intended region and whether its performance remains consistent during development, quality control and regulatory testing.

Cascade impaction is widely used for APSD measurement because it separates an aerosol into fractions by inertial impaction. When testing using an NGI, the aerosol passes through a series of stages, each designed to collect particles above a defined aerodynamic cut-off size, following a saw-tooth path. Each stage has progressively more and smaller nozzles, so air velocity rises and smaller particles are collected at each step. Particles that pass stage 7 are captured by the Micro-Orifice Collector (MOC). The drug deposited on each stage is then recovered and assayed, typically using a suitable analytical method, such as HPLC.
See the video below to understand how cascade impaction works.
How cascade impaction for APSD measurement works
NGI stage cut-off diameters at 15, 30, 60 and 100 L/min
In cascade impaction, each stage is defined by a cut-off diameter: the aerodynamic particle size at which particles are separated between one stage and the next. Particles with sufficient inertia are collected on that stage, while smaller particles remain in the airflow and continue to the later stages. Cut-off diameters determine how the delivered dose is divided into aerodynamic size fractions during APSD measurement.
For the NGI, stage cut-off diameters depend on the test flow rate. This means the effective size range collected by each stage changes with flow, making accurate flow selection and control essential for reliable, reproducible and method-relevant APSD data.
The table below shows the NGI stage cut-off diameters at commonly used inhaler testing flow rates.
| Stage | 15 L/min | 30 L/min | 60 L/min | 100 L/min |
|---|---|---|---|---|
| Stage 1 | 14.10 | 11.72 | 8.06 | 6.12 |
| Stage 2 | 8.61 | 6.40 | 4.46 | 3.42 |
| Stage 3 | 5.39 | 3.99 | 2.82 | 2.18 |
| Stage 4 | 3.30 | 2.30 | 1.66 | 1.31 |
| Stage 5 | 2.08 | 1.36 | 0.94 | 0.72 |
| Stage 6 | 1.36 | 0.83 | 0.55 | 0.40 |
| Stage 7 | 0.98 | 0.54 | 0.34 | 0.24 |
Cut-off diameters at other flow rates between 30 and 100 L/min can be calculated using the equation given in Ph. Eur. 2.9.18 and USP <601>.
Good NGI testing practice
Reliable APSD measurement depends on a well-controlled and maintained NGI method. Good practice helps reduce variability, protect data integrity and maintain confidence in routine results.
System integrity and flow control
Leak testing and accurate volumetric flow control are essential, particularly because NGI stage cut-off diameters are flow-rate dependent.
Cleaning and collection surfaces
Regular cleaning helps prevent residue, contamination and carryover, while appropriate surface coating can help minimise particle bounce and re-entrainment.
Performance assurance
Routine stage mensuration confirms that critical NGI dimensions remain within specification, while environmental conditions such as temperature, humidity and electrostatics should be controlled where relevant.
Copley supports ongoing NGI performance with stage mensuration, servicing, automation and workflow optimisation.
Which inhaler types can be tested with the NGI?
The NGI is used for APSD testing across all major orally inhaled and nasal drug product types:
- Metered-dose inhalers (MDIs) and breath-actuated inhalers
- Dry powder inhalers (DPIs), with a preseparator
- Soft mist inhalers (SMIs)
- Nebulisers, at 15 L/min with the NGI Cooler™
- Nasal sprays, aerosols and powders
- Spacers and valved holding chambers (VHCs)
NGI resources and downloads
Next Generation Impactor FAQs
NGI stands for Next Generation Impactor. It is a seven-stage cascade impactor with a Micro-Orifice Collector, used to measure the aerodynamic particle size distribution of inhaled and nasal drug products. It was first described as the “next generation pharmaceutical impactor” and is sometimes referred to as the “new generation impactor”.
The Next Generation Impactor NGI is used to determine the Aerodynamic Particle Size Distribution (APSD) of orally inhaled and nasal drug products. It separates the dose by aerodynamic particle size, allowing analysts to quantify how much active pharmaceutical ingredient is collected at each stage of the impactor.
APSD is a critical quality attribute for inhaled products, as particle size influences deposition behaviour within the respiratory tract. Cascade impaction provides a standardised in vitro method for assessing product performance during development, quality control and stability testing.
The NGI draws the sample through a series of precision-engineered nozzles and collection stages. Larger particles impact on the earlier stages, while progressively smaller particles remain entrained in the airflow and are collected on later stages. This enables analysts to generate a particle size distribution profile across the respirable range.
The NGI is calibrated for use from 30 to 100 L/min, with 15 L/min used for nebuliser testing. MDIs are typically tested at 30 L/min, while DPIs are tested at the flow rate that gives a 4 kPa pressure drop across the device, up to 100 L/min. Stage cut-off diameters change with flow rate (see the table above).
The Micro-Orifice Collector, or MOC, is the final collection stage of the NGI. It is designed to collect the finest particles that pass through Stage 7, allowing these particles to be recovered and analysed in a similar way to material collected from the other stage cups.
Yes. The NGI’s precision-engineered design, removable cup tray, low inter-stage losses and compatibility with automation make it well suited to routine APSD measurement in quality control laboratories, provided the method is appropriately developed, validated and maintained.
A preseparator is typically used for dry powder inhaler testing. It helps capture large, non-inhalable powder agglomerates before they enter the impactor, reducing the risk of overloading the upper stages and helping to maintain the integrity of the APSD measurement data.
The NGI is supported by archival calibration data, while routine performance assurance is managed through stage mensuration. We recommend annual NGI stage mensuration to confirm that impactor dimensions remain within specification.
Find out more about NGI stage mensuration and servicing.
After testing, deposited drug is recovered from the relevant components of the test set-up, including the induction port, preseparator where used, collection cups and MOC. A suitable recovery solvent is typically added to each collection surface before the samples are prepared for chemical analysis.
Please see our automated solutions for APSD measurement and our new Sample Recovery System™ SRS 100i to learn more about drug recovery.
Both the NGI and Andersen Cascade Impactor (ACI) are used for APSD measurement, but they differ in design and workflow. The ACI is a vertically stacked impactor, while the NGI uses a horizontal seven-stage design with removable collection cups and a Micro-Orifice Collector (MOC). This makes the NGI particularly well suited to pharmaceutical inhaler testing workflows where efficient sample recovery, reproducibility, throughput and automation compatibility are important.
The ACI has eight stages and is calibrated at 28.3 L/min, and remains widely used, particularly for MDIs and established methods. Find out more about the Andersen Cascade Impactor (ACI).
Yes. Key stages of the NGI workflow can be semi-automated to improve reproducibility and throughput, including impaction surface coating, sample recovery and data handling. Copley offers dedicated automation solutions such as the Sample Recovery System™ SRS 100i and Impactor Genie™ IG 200i.
Related Applications
We also offer a range of equipment to support APSD measurement, using the NGI
Related Services
Training
We offer a range of training courses, presentations and seminars covering a wide range of topics.
Find Out More
Servicing
A comprehensive range of both in-house and on-site product servicing options are available
Find Out More
Support
Our team of experienced technicians and engineers are on hand to help and advise
Find Out More